The G-16 Elite - Temperature and Temperature/Humidity

Temperature and Temperature/Humidity Test Chambers
The G-16 Elite

CE Certification on all Elite Models

Russells G-16 Elite Series temperature and temperature/humidity chambers represent one of our pre-engineered environmental chambers. This model has a chamber volume in 16 cubic feet or 439 liters. Click for a complete overview highlighting the features and benefits of our Elite Series as well as other standard model sizes to meet your simulation needs in testing extremes in cold and heat on your products. This reach-in chamber model is available with cryogenic cooling!

A Russells application engineer can assist you in selecting the laboratory chamber configuration, options, and size that's best for you, custom designed to your exact specifications.

Temperature and Temperature/Humidity Test Chambers – The G-16
G (B/D) - 16
Printable PDF - G-16 Temperature - Humidity Chambers
 
Specifications:
   Chamber Model --- *denotes, (B = temperature, D = humidity) G*-16-3 G*-16-3-3
   Chamber Volume (cu ft) / (liters) 16 / 439
Interior: W x D x H (in) 30 x 30 x 30
(cm) 76 x 76 x 76
Exterior: W x D x H (in) 40 x 60 x 85
(cm) 102 x 153 x 216
   Shipping Weight: (lb) / (kg) 1,000 / 450 1,200 / 550
   Compressor(s): Hi stage (hp) / lo stage (hp) 3 / N.A. 3 / 3
   Chamber Heaters: (kw) 6 6
   Humidity Heaters: (GD models only) (kw) 2 2
   Power Requirements: Voltage / Phase / Frequency 230 / 3/ 60 230 / 3 / 60
Load in (kva) 15 19
   Performance:  
Temperature Range : ˚F -22  to  350 -90  to 350
˚C -30  to 177 -68  to  177
   Pull-Down (chamber empty) In Minutes
75˚F / 23˚C to 0˚F / -17˚C 35 8
to -40˚F / -40˚C N.A. 12
to -65˚F / -54˚C N.A. 15
to -85˚F / -65˚C N.A. 20
350˚F / 177˚C to 75˚F / 23˚C   20
   Heat-Up (chamber empty) In Minutes
75˚F / 23˚C to 240˚F / 116˚C 18 18
to 350˚F / 177˚C 25 25
-90˚F / -68˚C to 75˚F / 23˚C N.A. 12
   Live load capacity (in watts) In Watts
   (steady state) 0˚F / -17˚C 1,000 2,000
-40˚F / -40˚C N.A. 1,500
-65˚F / -54˚C N.A. 1,000
Humidity range (GD) Models:    10% RH to 95% RH in the range bounded by a 185˚F / 85˚C maximum chamber    temperature and a 40˚F / 4˚C minimum dewpoint.
Request-A-Quote - Temperature/Humidity Test Chambers