The G-8 - Temperature, Temperature Humidity

Laboratory Test Chambers
for Temperature and Temperature/Humidity Testing
The G-8 Elite

CE Certification on all Elite Models

Russells G-8 Elite Series temperature and temperature/humidity chambers represent one of our pre-engineered environmental chambers. This model has a chamber volume in 8 cubic feet or 226 liters. Click for a complete overview highlighting the features and benefits of our Elite Series as well as other standard model sizes to meet your needs. This reach-in chamber model is available with cryogenic cooling!

A Russells application engineer can assist you in selecting the laboratory chamber configuration, options, and size that's best for you, custom designed to your exact specifications.

Laboratory Test Chambers for Temperature and Temperature/Humidity Testing
G (B/D) - 8
Printable PDF - G-8 Laboratory Test Chambers
 
 
Specifications:
   Laboratory Chamber Model --- *denotes, ( B = temp, D = hum ) G*-8-105 G*-8-105-105
  Chamber Volume (cu ft) / (liters) 8 / 226
  Interior: W x D x H (in) 24 x 24 x 24
(cm) 61 x 61 x 61
  Exterior: W x D x H (in) 32 x 50 x 77
(cm) 82 x 127 x 196
Shipping Weight :   (lb) / (kg) 650 / 300 750 / 340
Compressor(s) :   Hi stage (hp) / Lo stage (hp) 1.5 / N.A. 1.5 / 1.5
Chamber Heaters :   (kw) 2 2
Humidity Heaters :   (GD models only) (kw) 2 2
  Power Requirements: Voltage / Phase / Frequency 230/1 or 3/60 230/1 or 3/60
Load in (kva) 8 10
  Performance:

 

Temperature Range : ˚F -22  to  350 -90  to 350
˚C -30  to 177 -68  to  177
  Pull-Down (chamber empty) In Minutes
75˚F / 23˚C to 0˚F / -17˚C 15 10
to -40˚F / -40˚C N.A. 15
to -65˚F / -54˚C N.A. 25
to -85˚F / -65˚C N.A. 30
350˚F / 177˚C to 75˚F / 23˚C   25
  Heat-Up (chamber empty) In Minutes
75˚F / 23˚C to 240˚F / 116˚C 20 20
to 350˚F / 177˚C 40 40
-90˚F / -68˚C to 75˚F / 23˚C N.A. 25
  Live load capacity (in watts) In Watts
(Steady State) 0˚F / -17˚C 500 750
-40˚F / -40˚C N.A. 600
-65˚F / -54˚C N.A. 450
Humidity range (GD) Models:   10% RH to 95% RH in the range bounded by a 185˚F / 85˚C maximum chamber   temperature and a 40˚F / 4˚C minimum dewpoint.
Laboratory Test Chambers, Lab Test Chamber Testing Extreme Heat and Cold for Your Products