Environmental Chambers that Give you the Technical Edge
CE Certification on all Elite Models
Russells G-64 Elite Series temperature and temperature/humidity chambers represent one of our pre-engineered environmental chambers. This model has a chamber volume in 64 cubic feet or 1,796 liters. Click for a complete overview highlighting the features and benefits of our Elite Series as well as other standard model sizes to meet your simulation needs in testing extremes in cold and heat on your products. This reach-in chamber model is available with cyrogenic cooling!
A Russells application engineer can assist you in selecting the temperature/humidity simulation chamber configuration, options, and size that’s best for you, custom designed to your exact specifications.
Chamber Model — *denotes,( B = temp, D = hum ) | G*-64-5 | G*-64-5-5 | |||
Chamber Volume (cu ft) / (liters) | 64 / 1,796 | ||||
Interior: W x D x H | (in) | 48 x 48 x 48 | |||
(cm) | 122 x 122 x 122 | ||||
Exterior: W x D x H | (in) | 56 x 78 x 101 | |||
(cm) | 142 x 198 x 257 | ||||
Shipping Weight : | (lb) / (kg) | 2,200 / 1,000 | 2,500 / 1,136 | ||
Compressor(s) : | Hi stage (hp) / lo stage (hp) | 5 / N.A. | 5 / 5 | ||
Chamber Heaters : | (kw) | 6 | 6 | ||
Humidity Heaters : | (GD models only) (kw) | 2 | 2 | ||
Power Requirements: | Voltage / Phase / Frequency | 460 / 3/ 60 | 460 / 3 / 60 | ||
Load in (kva) | 20 | 30 | |||
Performance: | |||||
Temperature Range | ˚F | -22 to 350 | -90 to 350 | ||
˚C | -30 to 177 | -68 to 177 | |||
Pull-Down (chamber empty) | In Minutes | In Minutes | |||
75˚F / 23˚C | 0˚F / -17˚C | 35 | 15 | ||
-40˚F / -40˚C | N.A. | 20 | |||
-65˚F / -54˚C | N.A. | 30 | |||
-85˚F / -65˚C | N.A. | 45 | |||
350˚F / 177˚C | 75˚F / 23˚C | 35 | 35 | ||
Heat-Up (chamber empty) | In Minutes | In Minutes | |||
75˚F / 23˚C | 240˚F / 116˚C | 25 | 25 | ||
350˚F / 177˚C | 40 | 40 | |||
-90˚F / -68˚C | 75˚F / 23˚C | N.A. | N.A. | ||
Live load capacity (in watts) | In Watts | ||||
(steady state) | 0˚F / -17˚C | 3,000 | 4,000 | ||
-40˚F / -40˚C | NA | 2,500 | |||
-65˚F / -54˚C | NA | 1,500 | |||
Humidity range (GD) Models: | 10% RH to 95% RH in the range bounded by a 185˚F / 85˚C maximum chamber temperature and a 40˚F / 4˚C minimum dewpoint. |