Environmental Chambers that Give you the Technical Edge

Request A Quote

The G-8 – Temperature, Temperature Humidity

  • Laboratory Test Chambers for Temperature and Temperature/Humidity Testing The G-8 Elite

    CE Certification on all Elite Models

    Russells G-8 Elite Series temperature and temperature/humidity chambers represent one of our pre-engineered environmental chambers. This model has a chamber volume in 8 cubic feet or 226 liters. Click for a complete overview highlighting the features and benefits of our Elite Series as well as other standard model sizes to meet your needs. This reach-in chamber model is available with cryogenic cooling!

    A Russells application engineer can assist you in selecting the laboratory chamber configuration, options, and size that’s best for you, custom designed to your exact specifications.

Specifications

Laboratory Chamber Model — *denotes, ( B = temp, D = hum ) G*-8-105 G*-8-105-105
  Chamber Volume (cu ft) / (liters) 8 / 226
  Interior: W x D x H (in) 24 x 24 x 24
(cm) 61 x 61 x 61
  Exterior: W x D x H (in) 32 x 50 x 77
(cm) 82 x 127 x 196
Shipping Weight :   (lb) / (kg) 650 / 300 750 / 340
Compressor(s) :   Hi stage (hp) / Lo stage (hp) 1.5 / N.A. 1.5 / 1.5
Chamber Heaters :   (kw) 2 2
Humidity Heaters :   (GD models only) (kw) 2 2
  Power Requirements: Voltage / Phase / Frequency 230/1 or 3/60 230/1 or 3/60
Load in (kva) 8 10
  Performance:

 

Temperature Range : ˚F -22  to  350 -90  to 350
˚C -30  to 177 -68  to  177
  Pull-Down (chamber empty) In Minutes
75˚F / 23˚C to 0˚F / -17˚C 15 10
to -40˚F / -40˚C N.A. 15
to -65˚F / -54˚C N.A. 25
to -85˚F / -65˚C N.A. 30
350˚F / 177˚C to 75˚F / 23˚C   25
Heat-Up (chamber empty) In Minutes
75˚F / 23˚C to 240˚F / 116˚C 20 20
to 350˚F / 177˚C 40 40
-90˚F / -68˚C to 75˚F / 23˚C N.A. 25
Live load capacity (in watts) In Watts
(Steady State) 0˚F / -17˚C 500 750
-40˚F / -40˚C N.A. 600
-65˚F / -54˚C N.A. 450
Humidity range (GD) Models:   10% RH to 95% RH in the range bounded by a 185˚F / 85˚C maximum chamber   temperature and a 40˚F / 4˚C minimum dewpoint.